Items |
Failure |
Handling Method |
|
1 |
C1 Compressor over-current tripping. |
Check whether the electromagnetic valve switch overload protector in distribution room is tripping. Please press the reset button again. |
|
2 |
C1 Compressor high voltage tripping. |
Open the cover below the body and check whether the high voltage protector switch is tripping. Please press the reset button again. |
|
3 |
C2 over-current tripping. |
Check whether the electromagnetic valve switch overload protector in distribution room is tripping. Please press the reset button again. |
|
4 |
C2 high voltage tripping. |
Open the cover below the body and check whether the high voltage protector switch is tripping. Please press the reset button again. |
|
5 |
Abnormal temperature control. |
Check whether the fan inside is running. If the fan motor is burn out please inform us to replace it Check whether the humidifier over-temperature protector is set improperly. If the protector isn’t in proper position please reset it. |
|
6 |
Abnormal power supply. |
*Please check if the power supply phase loss or inversion. *Please check wire is disconnected or not. |
|
7 |
Inner Chamber over-temperature |
Check whether the over-temperature protector is damaged or cannot control. Please inform us to repair. |